Agilent Technologies

Big Universal Trap

Installation Sheet

Agilent Gas Purifiers

Maximum operating pressure is 250 PSIG.
Maximum operating temperature is 100 °C.
Maximum flow rate is 8 liters per minute.

Installing the Big Universal Trap

Before installing the trap, make sure the system is free of leaks, the GC inlet and MSD 
vacuum chambers are clean (if using mass spectrometry detection), and the system is 
generally in good working order.

1

Shut down your GC/MSD or reduce the GC heated zone temperatures.

2

Set the gas source supply pressure to 10–15 psi and maintain flow in the gas source 
line before disconnecting it from the inlet of the old trap.

3

Remove the protective nut and plug from the 

INLET

 end of the new trap. 

DO NOT

 

open the plug on the 

OUTLET

 end.

4

Immediately attach the new trap to the gas source tubing using one of the included 
ferrule sets.

5

Insert the tubing through the nut and ferrule set until the tubing rests firmly 
against the shoulder in the fitting.

6

Finger-tighten the nut and then use a wrench to tighten the nut 3/4 turn for 
1/8-inch tubing and 1-1/4 turn for 1/4-inch tubing (see 

Figure 1

).

Part number

Fitting

Gas

RMSH–2

1/8-inch

Helium

RMSH–4

1/4-inch

Helium

RMSN–2

1/8-inch

Nitrogen

RMSN–4

1/4-inch

Nitrogen

RMSHY–2

1/8-inch

Hydrogen

RMSHY–4

1/4-inch

Hydrogen

C A U T I O N

Failure to immediately connect the trap may cause contamination of the adsorbents inside, 
which will reduce its adsorption capacity and contribute to elevated levels of contaminants 
observed by the detector. This may result in the need for additional purge time.

Figure 1

Outlet fitting closed

OUTLET

Outlet fitting closed

Inlet fitting

99.997 + % He, N

2

, or H

2