Wait for 3 minutes to stabilize the gas pressure inside the new trap. Increase the 
gas source pressure to 60 psi.

8

Open the outlet fitting and purge the trap for 3 minutes (see 

Figure 2

).

9

Adjust the gas source pressure to a normal working level. Connect the outlet fitting 
of the trap to the instrument tubing as described in steps 

5

 and 

6

. Wait for at least 

3 minutes to purge out any air that may have entered the instrument tubing line. 
The trap is now ready for use (see 

Figure 3

).

For the latest detailed MSDS information, please visit our website at 
www.agilent.com/chem/msds.

© Agilent Technologies, Inc. 2002

No part of this document may be reproduced in any form or by any means (including electronic storage and retrieval or trans-
lation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by 
United States and international copyright laws.

First edition, May 2002

Printed in USA

Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA

Figure 2

Outlet fitting open

OUTLET

Outlet fitting open

Inlet fitting

99.997 + % He, N

2

, or H

2

N O T E

Most instruments (such as GCs or GC-MSDs) require a waiting time for the system to stabilize 
and self-clean before use when the gas source or gas trap is changed. This waiting time is often 
2 hours or longer. Results from analyses performed too soon after changing the trap may not be 
reproducible. The longer the wait time the more stable the system will be, thus protecting the 
instrument and the column when changing gases or the gas trap.

Figure 3

Completed installation

OUTLET

Instrument

5184-3513