6
Installation Note
N9039-90004
Option BAB, 3.5 mm RF Connectors Option 518, Type-N RF Connectors Installation Kit
Installation Procedure
CAUTION
If the instrument is placed on its face during any of the following
procedures, be sure to use a soft surface or soft cloth to avoid damage to
the front panel, keys, or input connector.
1. Disconnect the instrument from ac power.
2. Refer to
. Using the T-20 driver, remove the four screws (two on each side)
(1)
that attach the handle strap
(2)
on each side of the instrument.
3. Using the T-20 driver, remove the four screws (including washers)
(3)
that hold the
rear feet
(4)
in place.
4. Pull the instrument cover
(5)
off towards the rear of the instrument.
Figure 1
Instrument Outer Case Removal