![Agilent Technologies N9010A Скачать руководство пользователя страница 61](http://html1.mh-extra.com/html/agilent-technologies/n9010a/n9010a_manual_2868178061.webp)
Chapter 8
61
Power Statistics CCDF Measurements
Measurement Procedure
Measurement Procedure
Step 1.
Press
Mode
,
TD-SCDMA with HSPA/8PSK
to enable the TD-SCDMA with
HSPA/8PSK mode measurements.
Step 2.
Press
Mode Preset
to preset the analyzer.
Step 3.
Select the signal inputs for the measurement.
For BBIQ measurements this will be the I or I input and the Q or Q
input. Use the procedure in “Using Option BBA Baseband I/Q Inputs.“
Step 4.
Press
Input/Output
,
External Gain
,
BTS
and enter the number and
−
dB
according to the external attenuator connected to the analyzer.
Step 5.
Press
Mode Setup
,
Radio
,
Device
to toggle the device to
BTS
.
Step 6.
Press
Meas
,
Power Stat CCDF
to initiate the power statistics CCDF
measurement.
Step 7.
Press
Trigger
to select the appropriate trigger source
External 1
.
If no measurement is performed and only a symbol of yellow ‘*’ displays
on the top right of the screen, you need to check the trigger settings.
You need the correct trigger signal to make the CCDF measurement.
Step 8.
Press
FREQ Channel
,
2
,
GHz
to set the center frequency to 2 GHz.
Step 9.
Press
Meas Setup
to check
the keys available to change the measurement
parameters from the default condition.
For more information about changing measurement parameters, see
N9079A TD-SCDMA with HSPA/8PSK User’s and Programmer’s
Reference
.
If you have a problem, and get an error message, see the “
Instrument
Messages Guide
”.
Содержание N9010A
Страница 4: ......
Страница 7: ...Contents 7 Waveform Signal Envelope Views of I only or Q only 106...
Страница 8: ...8 Contents...
Страница 12: ...12 Chapter1 Making TD SCDMA with HSPA 8PSK Measurements Setting Up and Making a Measurement...
Страница 25: ...Chapter 3 25 ACP Measurements Measurement Procedure Messages Guide...
Страница 28: ...28 Chapter3 ACP Measurements Measurement Results...
Страница 44: ...44 Chapter5 Spectrum Emission Mask Measurements Troubleshooting Hints...
Страница 64: ...64 Chapter8 Power Statistics CCDF Measurements Troubleshooting Hints...
Страница 94: ...94 Chapter12 Monitor Spectrum Measurements Measurement Results...
Страница 95: ...95 13 Using Option BBA Baseband I Q Inputs...
Страница 103: ...Chapter 14 103 Baseband I Q Inputs Option BBA Measurement Concepts Why Make Measurements at Baseband...