![Agilent Technologies N9010A Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/agilent-technologies/n9010a/n9010a_manual_2868178019.webp)
Chapter 2
19
Transmit Power Measurements
Measurement Results
shows the Transmit Power measured by burst width.
Full Burst Width and Measured Width (Burst Width) are displayed in the text
window, as well as the Threshold levels used to calculate the burst center. When
the default Burst Width is inappropriate to perform a measurement, it’s
recommended to increase the Meas Interval and Threshold Lvl or to manually
enter the desired Burst Width value.
Figure 2-3
Transmit Power - Measured Burst Width
shows the Transmit Power above threshold level.
Absolute Amplitude Threshold and Relative Amplitude Threshold are displayed.
The relative threshold is defined as relative to the peak power.
Figure 2-4
Transmit Power - Above Threshold Lvl
Содержание N9010A
Страница 4: ......
Страница 7: ...Contents 7 Waveform Signal Envelope Views of I only or Q only 106...
Страница 8: ...8 Contents...
Страница 12: ...12 Chapter1 Making TD SCDMA with HSPA 8PSK Measurements Setting Up and Making a Measurement...
Страница 25: ...Chapter 3 25 ACP Measurements Measurement Procedure Messages Guide...
Страница 28: ...28 Chapter3 ACP Measurements Measurement Results...
Страница 44: ...44 Chapter5 Spectrum Emission Mask Measurements Troubleshooting Hints...
Страница 64: ...64 Chapter8 Power Statistics CCDF Measurements Troubleshooting Hints...
Страница 94: ...94 Chapter12 Monitor Spectrum Measurements Measurement Results...
Страница 95: ...95 13 Using Option BBA Baseband I Q Inputs...
Страница 103: ...Chapter 14 103 Baseband I Q Inputs Option BBA Measurement Concepts Why Make Measurements at Baseband...