3-38
Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
Adjustments
N5242A
Source Adjustment
The source calibration is used to adjust your network analyzer for a flat source power
across its full frequency range. There are differences between each test port; therefore, an
adjustment is required for each port.
Equipment Used for the Source Adjustment
Procedure
1. Connect the equipment as shown in
. Connect a GPIB cable between the
network analyzer and the power meter.
Figure 3-14 Equipment Setup for the Source Adjustment
2. .Press UTILITY
, then
, then
, then click
Source
Adjustment
.
3. Ensure the GPIB settings are correct.
4. Follow the instructions and prompts as they are displayed.
Equipment Type
Model or
Part Number
Alternate Model or
Part Number
Power meter
E4418B/E4419B
E4418A/E4419A
Power sensor, 3.5 mm
E4413A
8485A
Adapter, 3.5 mm (f) to 3.5 mm (f)
83059B
85052-60012
System
Service
Adjustments
Содержание N5242A
Страница 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Страница 33: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Страница 75: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Страница 119: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Страница 153: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Страница 237: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Страница 303: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...
Страница 342: ...Index 22 Service Guide N5242 90001 Index...