3-26
Service Guide N5242-90001
Tests and Adjustments
PNA Series Microwave Network Analyzers
System Verification
N5242A
Interpreting the Verification Results
shows an example of typical verification results with
Print Tabular Data
selected in the
Printer Output
area of the
System Verification
dialog box.
At the top of the printed output is the name of the device, the serial number of the device,
and the date tested.
Each S-parameter measurement result is printed with frequency tested, lower and upper
limit lines, the measured data, and the result of the test.
Figure 3-11 Example of Printed Tabular Verification Results