Service Guide N5242-90001
2-13
PNA Series Microwave Network Analyzers
General Product Information
N5241A/42A
Required Service Test Equipment
Required Service Test Equipment (Cont’d)
Equipment
a
a. Unless specified otherwise, equipment listed is required for all analyzer models.
Critical Specifications
Recommended
Model or Part
Number
Alternate Model
or Part Number
Use
b
b. P = Performance tests, A = Adjustments, T = Troubleshooting, R = Repair, V = System verification
Calibration and Verification Kits
3.5 mm calibration kit
Freq: 10 MHz to 26.5 GHz
85052B
85052D
P,T,V
3.5 mm verification kit
Freq: 10 MHz to 26.5 GHz
85053B
None
V
Cables
BNC cable (2 required)
50
Ω
, length
≥
60 cm
8120-1839
None
A
3.5 mm RF cable (Qty 2)
50
Ω
, length
≥
60 cm
85131C
85131E
P,A,V
GPIB cable
N/A
10833A/B/C/D
None
P,A
Adapters
3.5 mm (f) to 3.5 mm (f)
Return Loss:
≥
32 dB
83059B
85052-60012
c
c. Included in the 85052B/D calibration kits.
P,A,T
3.5 mm (f) to type-N (m)
Return Loss:
≥
28 dB
1250-1743
None
P,A,T
Attenuators
3.5 mm (m,f), 10-dB fixed
attenuator
Accuracy: ± 0.5 dB
Freq: 10 MHz to 26.5 GHz
8493C
Option 010
None
P
3.5 mm (m,f), 20-dB fixed
attenuator
Accuracy: ± 0.5 dB
Freq: 10 MHz to 26.5 GHz
8493C
Option 020
None
P
Содержание N5241A
Страница 11: ...Service Guide N5242 90001 1 1 1 Safety and Regulatory Information...
Страница 19: ...Service Guide N5242 90001 2 1 2 General Product Information...
Страница 35: ...Service Guide N5242 90001 3 1 3 Tests and Adjustments...
Страница 81: ...Service Guide N5242 90001 4 1 4 Troubleshooting...
Страница 125: ...Service Guide N5242 90001 5 1 5 Theory of Operation...
Страница 159: ...Service Guide N5242 90001 6 1 6 Replaceable Parts...
Страница 243: ...Service Guide N5242 90001 7 1 7 Repair and Replacement Procedures...
Страница 313: ...Service Guide N5242 90001 8 1 8 General Purpose Maintenance Procedures...
Страница 354: ...Index 22 Service Guide N5242 90001 Index...