N2780/81/82/83 User’s and Service Guide
43
test procedure. While the test system is warming up,
clean the magnetic contacts on the probe jaw.
2
Press the
DEMAG
button on the probe terminator.
3
Configure the DMM to measure AC Volts.
4
Clamp the probe around the coil and ensure that the
probe is as perpendicular as possible to the coils of wire.
5
Set up the signal generator as follows:
•
Wave shape: Sine
•
Frequency: 50 Hz
01
01
COM
COM
SENSE
!
Signal Generator
DMM
Probe
Power
Supply
Current Probe
Test Coil
Adapter