N2780/81/82/83 User’s and Service Guide
31
Measurement Procedure
Preparations for Measurement
1
Have the N2779A power supply and oscilloscope ready for
waveform measurement ready.
2
Turn the power switch off and connect the power cord.
3
Connect the power plug of the current probe to the power
receptacle of the N2779A.
4
Turn the N2779A power switch on, and check that the
front panel power indicator lights.
Demagnetization and Zero Adjustment
1
With the oscilloscope input at ground, adjust the trace to
the zero position.
2
Set the input coupling of the oscilloscope to DC.
C A U T I O N
Before turning on the power, make sure that the voltage of
the power supply being used matches the supply voltage
indicated on the rear panel of the N2779A.
N O T E
The output of this unit is terminated internally. Use a high
impedance input to the measuring instrument. Accurate
measurements are not possible when the input impedance of
the oscilloscope is set to 50
. Besure to set the input
impedance to 1 M
before making measurements. If the
oscilloscope input impedance cannot be set to 1 M
then the
Agilent E2697A impedance adaptor can be purchased.