ERC Protocol
M4735A Heartstream XL Defibrillator/Monitor
3-19
3
If CPR is needed, press
. While paused, a timer indicates the elapsed
time and the total duration of the Pause state, as shown:
)LJXUH'LVSOD\LQ3DXVHG6WDWH
The total pause duration depends on the event preceding the Pause state. If
you entered the Pause state:
l
at the completion of a shock series or shortly after a shock is deliv-
ered, the duration is equal to the
3RVW6KRFN&357LPHU
configuration set-
ting (the default setting is 60 seconds).
l
when no shock was advised, the duration is equal to the
16$7LPHU
configuration setting, where NSA is an acronym for No Shock
Advised (the default setting is 180 seconds).
3$86(
Shocks: 3
00:01:40
Paused
RESUME
ANALYZE
Timer: 35 / 60
Elapsed Time
Total Pause Duration
in Seconds
Содержание M4735A
Страница 1: ... JLOHQW 0 HDUWVWUHDP HILEULOODWRU 0RQLWRU 8VHU V XLGH ...
Страница 2: ......
Страница 3: ...User s Guide 0 HDUWVWUHDP HILEULOODWRU 0RQLWRU ...
Страница 6: ...vi ...
Страница 24: ...Documentation and Training 1 10 Introduction ...
Страница 40: ...Using a Data Card Optional 2 16 Getting Started ...
Страница 62: ...Troubleshooting 3 22 Defibrillating in AED Mode ...
Страница 96: ...Entering AED Mode 6 10 Defibrillating in Manual Mode ...
Страница 104: ...Disabling Sync Mode 7 8 Performing Synchronized Cardioversion ...
Страница 114: ...Troubleshooting 8 10 Pacing Optional ...
Страница 156: ...Disposing of the Heartstream XL 11 18 Maintaining the Heartstream XL ...
Страница 192: ...Electromagnetic Compatibility 13 28 Specifications Safety ...
Страница 201: ......
Страница 202: ... GLWLRQ 0 RS ULJKW JLOHQW 7HFKQRORJLHV QF 3ULQWHG LQ WKH 8 6 6HSWHPEHU ...