Specifications
13-10
Specifications & Safety
Auto Printing: The printer can be configured to automatically print on Mark,
Charge, Shock and Alarm.
Delayed Printing: The printer can be configured to run real time or with a 6
second delay.
Reports: The following can be printed: Event Summary, Configuration,
Extended Self Test, System Log, Battery Capacity Test, Shift/System Check.
Speed: 25 mm/s with an accuracy of ± 5%.
Amplitude Accuracy: ± 10% or ± 50 uV, whichever
is greater.
Paper Size: 50 mm by 30 m (100 ft).
Noninvasive Pacing
Waveform: Monophasic Truncated Exponential.
Current Pulse Amplitude: 10 mA to 200 mA (5 mA resolution); accuracy 10
mA to 50mA ± 5mA, 50mA - 200mA± 10%.
Pulse Width: 20 ms with ac0, -5 ms.
Rate: 30 ppm to 180 ppm (10 ppm increments); accuracy ± 1.5%.
Modes: Demand or Fixed Rate.
Refractory Period: 340 msec (30 to 80 ppm); 240 msec (90 to 180 ppm).
Содержание M4735A
Страница 1: ... JLOHQW 0 HDUWVWUHDP HILEULOODWRU 0RQLWRU 8VHU V XLGH ...
Страница 2: ......
Страница 3: ...User s Guide 0 HDUWVWUHDP HILEULOODWRU 0RQLWRU ...
Страница 6: ...vi ...
Страница 24: ...Documentation and Training 1 10 Introduction ...
Страница 40: ...Using a Data Card Optional 2 16 Getting Started ...
Страница 62: ...Troubleshooting 3 22 Defibrillating in AED Mode ...
Страница 96: ...Entering AED Mode 6 10 Defibrillating in Manual Mode ...
Страница 104: ...Disabling Sync Mode 7 8 Performing Synchronized Cardioversion ...
Страница 114: ...Troubleshooting 8 10 Pacing Optional ...
Страница 156: ...Disposing of the Heartstream XL 11 18 Maintaining the Heartstream XL ...
Страница 192: ...Electromagnetic Compatibility 13 28 Specifications Safety ...
Страница 201: ......
Страница 202: ... GLWLRQ 0 RS ULJKW JLOHQW 7HFKQRORJLHV QF 3ULQWHG LQ WKH 8 6 6HSWHPEHU ...