Compact Cold Cathode Gauge IMG-500
11
2
Technical Data
Measuring range (air, N
2
)
2×10
-9
… 1×10
-2
mbar
Accuracy
±30% (in the range
1×10
-8
… 1×10
-3
mbar)
Reproducibility
±5% (in the range
1×10
-8
… 1×10
-3
mbar)
Output signal (measuring signal)
Voltage range
0
… +10.5 V
Voltage vs. pressure
logarithmic , 1.0 V/decade
Error signal
<0.5 V no supply
Output impedance
2×10
Minimum loaded impedance
10 k
, short-circuit proof
Response time
p > 10
-6
mbar
(pressure dependent)
<10 ms
p = 10
-8
mbar
1000 ms
Identification gauge
100 k
, referenced to
supply common
Supply
DANGER
The gauge may only be connected to power sup-
plies, instruments or control devices that conform
to the requirements of a grounded protective extra-
low voltage (SELV). The connection to the gauge
has to be fused
1)
.
Voltage at the gauge
15
… 30 VDC
(ripple
1 V
pp
)
1)
Agilent controllers fulfill this requirement.