85420E/85460A High Frequency Filter Assembly Replacement Kit 5
Remove the A9 HF Filter Assembly
1. Place the instrument with the top facing up on the work surface.
2. Remove eight screws on the front to access the components underneath
the top lid (1). Refer to Figure 0-2 on page -6.
3. Disconnect the five cables that go to the A9 HF Filter assembly (2).
Refer to Figure 0-3 on page -vi.
4. Disconnect the ribbon cable (3) from the HF filter assembly.
5. Remove the six screws that attach the HF filter housing to the front
strut (4).
6. Lift the housing out of the instrument.