HP/Agilent 85420E/85460AHigh Frequency Filter AssemblyReplacement Kit
Installation Note
Part number: 85460-90030Printed in USAApril 2000
Страница 1: ...HP Agilent 85420E 85460A High Frequency Filter Assembly Replacement Kit Installation Note Part number 85460 90030 Printed in USA April 2000...
Страница 2: ...nty of any kind with regard to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for er...
Страница 3: ...ervice Guide for the appropriate adjustment routines and performance verification tests that must be run after replacing the HF Filter Assembly Refer to Chapter 2 Making Adjustments Table 2 2 A9 High...
Страница 4: ...t from all voltage sources while it is being opened The power cord is connected to internal capacitors that may remain live for 5 seconds after disconnecting the plug from its power supply CAUTION To...
Страница 5: ...crews on the front to access the components underneath the top lid 1 Refer to Figure 0 2 on page 6 3 Disconnect the five cables that go to the A9 HF Filter assembly 2 Refer to Figure 0 3 on page vi 4...
Страница 6: ...6 85420E 85460A High Frequency Filter Assembly Replacement Kit Figure 0 2 A9 HF Filters Replacement Removing Lid Screws Figure 0 3 A9 HF Filters Replacement Lid Removal...
Страница 7: ...ge when replacing the instrument cover remember the following Place a soft cloth or towel between the work surface and the front frame Ensure that cables do not bind between the instrument cover and i...