Installation
Installing the Ferrite
Agilent G6697A Oil Free IDP-10 Dry Scroll Pump Installation Guide
21
Installing the Ferrite
Verify that the collision cell signal cable has a ferrite installed on it. This applies to
the 7000 and 7010 Series GC/MS.
1
Open the analyzer door cover to view the wiring harness and collision cell
board as shown in
2
Look at the wiring harness and the collision cell signal cable to see if it has a
ferrite on it already.
Figure 12.
Wiring harness and collision cell board