Operation
4
Installation and Operation Guide
51
In this case, port 1 is plugged and the column is connected to port 2. The two
detector restrictors are connected to ports 3 and 4. Calculations for a two-way
splitter are similar to those for the three-way. See the manual for the G3180B
Micro-Fluidic Effluent Splitter with Makeup Gas provided on the CD that came
with this product for setting up a two-way split.
Two columns in/Two detectors out
shows the configuration for a two-column combiner/two-way
splitter.
This configuration could be used in a system that allows injection into either
inlet 1 or inlet 2.
Figure 26
Two columns in/two detectors out
Aux EPC in
Det 2 out
Det 1 out
Column 2 in
Column 1 in
1
2
3
4