Operation
Select restrictors
Start up the spreadsheet "splitter3_calc_rev1.xls" in Excel. Select the “3-Way
(Makeup)” worksheet. We will choose to have 4 mL/min go to the MSD initially.
This flow is acceptable with a performance turbo system. Fill in the input
column as shown (
) with the MSD assumed to be Detector 1, the FPD
as Detector 2, and the ECD as Detector 3.
Figure 18
The Effluent Splitter calculator