1.
Loosen the four (4) Pozidrive screws holding the two attaching
fingers on the G1926A Bar Code Reader (BCR). These screws are
concealed by the cover, but there are access holes over each screw
position. See Figure 1.
2.
Loosen the screws just enough, about ¼ turn, to allow the fingers to
be adjusted easily but secure enough to hold the position until the
screws can be re-tightened.
Figure 1.
Bar code reader
Installing the G1926A Bar
Code Reader on the G2614A
Sampler Tray
Agilent 6890 Gas Chromatograph