FID Maintenance Kit Instructions
Install the Cleaned or Replacement Jet
If the jet is new, tighten the jet 1/6-turn past finger tight.
Install the Replacement or Cleaned Collector Assembly
1
Carefully lower the new or cleaned and reassembled
collector assembly straight down on to the FID base – the
interconnect spring will move into the correct position.
2
Secure it firmly with the three T20 Torx screws.
CAUTION
If you are re-using the old jet be very careful not to over-tighten the
jet to avoid breaking it off in the FID Base!