10
FID Maintenance Kit Instructions
Reassembling the FID collector assembly
1
Insert the first PTFE insulator from the kit into the collector
housing.
2
Insert the cleaned collector through the PTFE Insulator –
The long end goes down!
3
Place the second insulator over the collector.
Long end
Orientation of the collector and
insulators in the FID