Introduction to the Well-plate Sampler
6
1100 Series WS MWS Reference Manual
213
Electrical Connections
WA R N I N G
Never use cables other than the ones supplied by Agilent Technologies to ensure
proper functionality and compliance with safety or EMC regulations.
Figure 36
Well-plate Sampler Electrical Connections
DC-Out
RS232C
CAN-bus
Remote
Vial number output
Relay
contacts
CAN cable to next module
CAN cable to previous module
Control of Thermostat
WA R N I N G
DO NOT disconnect or reconnect the autosampler to ALS thermostat cable when the
power cords are connected to either of the two modules. This will damage the
electronics of the modules.