114
1100 Series WS MWS Reference Manual
4
Repairing the Sampler
Removing the Needle Carrier Assembly
1
In the user interface start the maintenance mode and select the “Change
Needle Carrier” function.
2
Open the front door and remove the side door.
3
Remove the plate tray from the tray base.
4
Push the silicon safety tube over the needle
5
Pinch the holder clamp, pull back and remove the needle assembly from the
needle carrier.
6
Unplug the flex board on the sample transport.
7
Unscrew the three holding hex screws with the 2 mm hex key.
8
Remove the needle carrier assy.
When required
When the needle carrier is defect
Tools required
2 mm hex key 8710-2438 (supplied in accessory kit)
Parts required
G1367-60010 Needle Carrier assembly
C A U T I O N
An uncovered needle is a risk of harm to the operator.
Содержание G1367A
Страница 1: ...s1 Agilent 1100 Series Well plate Sampler Micro Well plate Sampler Reference Manual...
Страница 4: ...4 1100 Series WS MWS Reference Manual...
Страница 42: ...42 1100 Series WS MWS Reference Manual 1 Installing the Sampler...
Страница 156: ...156 1100 Series WS MWS Reference Manual 4 Repairing the Sampler...
Страница 168: ...168 1100 Series WS MWS Reference Manual 5 Parts and Materials Figure 23 Injection Valve Assembly 1 2 4 3 6 5...
Страница 196: ...196 1100 Series WS MWS Reference Manual 5 Parts and Materials...
Страница 214: ...214 1100 Series WS MWS Reference Manual 6 Introduction to the Well plate Sampler...
Страница 240: ...240 1100 Series WS MWS Reference Manual 7 Theory of Operation...
Страница 276: ...276 1100 Series WS MWS Reference Manual 9 Specifications...
Страница 284: ...284 1100 Series WS MWS Reference Manual Index...
Страница 285: ......