Service Guide E8364-90001
3- 37
PNA Series Microwave Network Analyzers
Tests and Adjustments
E8362A, E8363A, E8364A
Performance Tests (Agilent N2721A Software Package)
Power Meter Accuracy Test
This test is intended for power meters used in testing the PNA series network analyzers.
The
requires the use of a power meter that has been calibrated to
a higher accuracy than the standard power meter.
Power meters with options G12 and H12 specify an improved instrumentation accuracy
over a limited power range. (These power meters do not contain unique hardware.) A
power meter may be returned to the factory to have one of these options added to an
existing power meter or to renew the calibration for one of these options.
This test procedure is an alternative to returning the power meter to the factory. When a
power meter passes this test, it is considered to be
calibrated
for the G12 or H12 option
even though it has not been returned to the factory.
This test procedure is available on the Internet as a separate document, Agilent Part
Number E8356-90051. The on-line document may have a more recent print date. Refer to
“Contacting Agilent” on page 2-9
for the Internet address (URL).
EPM Series Power Meters That Can Be Tested Using This Procedure
This procedure assumes that the recommended model number power meter is being tested.
The alternate model numbers can be tested but the necessary procedural steps may differ.
NOTE
It is recommended that the revision number for the power meter “Main
Firmware” be Ax.03.00 or higher. This applies to all power meter model
numbers listed above (both recommended and alternate).
Equipment Used for the Power Meter Accuracy Test
Recommended
Model Number
Alternate
Model Numbers
E4419B
E4418A/B, E4419A, EPM-441A, EPM-442A
Equipment Type
Recommended
Model or Part Number
Alternate
Model or Part Number
Range calibrator
11683A
None
Precision digital voltmeter
3458A
Any with the required accuracy and
resolution
a
a. Required accuracy and resolution at the following voltage levels:
14 mV input:
0.0100% accuracy
10 nV resolution
0.140 V input:
0.0050% accuracy
100 nV resolution
0.450 V input:
0.003% accuracy
100 nV resolution
Power sensor cable
8120-8319, 11730A
Any equivalent
Cable, BNC, 50
Ω
Any
Any
Adapter for connecting BNC
cable to DVM inputs
Any
Any
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Страница 302: ...A 16 Service Guide E8364 90001 Error Terms PNA Series Microwave Network Analyzers Error Term Data E8362A E8363A E8364A ...
Страница 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Страница 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Страница 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...