26
User’s Guide
E8362/3/4C Option H85
Example: Making High Power Measurements with Option H85
External Reference Mixer Switch Setup
1. In the
Trace/Chan
drop-down menu select
Channel
>
Hardware Setup
>
Test Set.
Figure 12
Trace/Chan
2. In the application window, select
External: flow through R1 Loop
.
Figure 13
R1 Input Path
3. Measure the output power at Test Port 1, using a power meter. Verify the power
measurement.
If you are measuring a highly reflective device, a high power isolators should be inserted
between the 20 dB coupler and CPLR THRU front panel ports to protect Port 1 Source.