28
Chapter 3
Operation Verification
Checking the Output Power
Perform the following procedure to check the signal generator output power:
1. Zero and calibrate the power sensor to the power meter as shown:
2. Connect the equipment as shown:
3. Press
Preset
.
4. Press
RF On/Off
. The
RF On
annunciator is now displayed.
5. Press
Mod On/Off
. The
Mod Off
annunciator is now displayed.
6. Press
Frequency
and enter the first frequency value listed in
Table 3-1
.
7. Press
Amplitude
and enter the amplitude value for your option and frequency value.
8. Measure the power level.
9. Repeat step
6
through
8
and measure the power levels for each remaining frequency listed in
Table 3-4
thru
Table 3-6
. Record any power levels not within the acceptable limits.