E7002, E7012, and E7014 Emergency Discharge Unit
1
Introduction
Safety
•
This equipment does not require any routine maintenance.
•
No parts of this equipment are designated as repairable
by service personnel.
•
There are no significant risks of reciprocal interference
posed by the presence of this equipment.
•
Do not situate this equipment in areas of high
electromagnetic interference such as gradient amplifiers.
•
This unit has been preset to a voltage range suitable for
the national supply. The unit has a selectable input
voltage: 100V~; 120V~; 230V~; 240V~. Fuses F1 and F2
must be rated to suit the input voltage:
•
If selected to 100V~ or 120V~, they must be rated at
T250mAL 250V.
•
If selected to 230V~ or 240V~, they must be rated at
T100mAL 250V.
WA R N I N G
Isolate or disconnect equipment from mains supply before opening this
unit for any reason. LIVE PARTS INSIDE.
This equipment must be connected to an earthed mains supply.
For continued protection against risk of fire or electric shock, replace
fuses with the same type, rating and UL listing.
WA R N I N G
This equipment is not suitable, and must not be used in areas with
flammable mixtures.
WA R N I N G
If replacing the back-up batteries, follow the instructions in
on page 25. The batteries are of a lead acid type and must be
disposed of at an appropriate recycling center.
C A U T I O N
This unit contains internal protection fuses. Follow the guidelines
below before replacement of these devices. Only suitably trained
personnel are permitted to remove covers on the equipment.
C A U T I O N
Do not position this equipment so that it is difficult to disconnect it
from the mains supply.
Содержание E7002
Страница 1: ...Agilent Technologies Agilent E7002 E7012 and E7014 Emergency Discharge Unit Reference Guide...
Страница 14: ...14 E7002 E7012 and E7014 Emergency Discharge Unit 2 Description and Specification...
Страница 22: ...22 E7002 E7012 and E7014 Emergency Discharge Unit 4 Troubleshooting...
Страница 24: ...24 E7002 E7012 and E7014 Emergency Discharge Unit 5 User Serviceable Parts...
Страница 31: ...Connections 7 E7002 E7012 and E7014 Emergency Discharge Unit 31 15 BULB...
Страница 32: ......
Страница 33: ...Agilent Technologies Agilent Technologies Inc Printed in USA Rev B Dec 2012 9100383500 9100383500...