![Agilent Technologies E5382A Скачать руководство пользователя страница 63](http://html.mh-extra.com/html/agilent-technologies/e5382a/e5382a_user-manual_2868081063.webp)
63
Operating the Probe
Combining grounds
Combining grounds
It is essential to ground every tip that is in use. For best performance at high
speeds, every tip should be grounded individually to ground in the system under
test. For convenience in connecting grounds, you can use the ground connector,
Agilent part number 16515-27601, to combine four probe tip grounds to connect
to one ground point in the system under test.
Using the 16515-27601 to combine grounds will have some negative impact on
performance due to coupling caused by common ground return currents. The
exact impact depends on the signals being tested and the configuration of the
test, so it is impossible to predict accurately. In general, the faster the rise time
of the signals under test, the greater the risk of coupling.
In no case should more than four tip grounds be combined through one 16515-
27601 to connect to ground in the system under test.
Содержание E5382A
Страница 4: ...4 Contents ...
Страница 5: ...1 General Information ...
Страница 13: ...2 Operating the Probe ...
Страница 64: ...64 Operating the Probe Combining grounds ...