Operating the Probe
5 cm Resistive Signal Lead and Solder-down Ground Lead
Other signal lead lengths may be used with these probes but a resistance value
needs to be determined from the following figure and a resistor of that value
needs to be placed as close as possible to the point being probed.
Optimum Damping Resistor Value Versus Signal Lead Length
If a resistor is not used, the response of the probe will be very peaked at high
frequencies. This will cause overshoot and ringing to be introduced in the step
response of waveforms with fast rise times. Use of this probe without a resistor
at the point being probed should be limited to measuring only waveforms with
slower rise times.
Length (cm)
Resista
nce (
Ω
)