Chapter 4
Source Key and Programming Commands
Normalize
59
So
ur
c
e
K
e
y a
nd Pr
ogram
min
g
Com
m
a
n
ds
4.5.2
Normalize
Enables or disables trace normalization.
Mode:
Spectrum Analysis
Key Path:
Source, Normalize
Remote Command:
:CALCulate:NTData[:STATe] ON|OFF|1|0
:CALCulate:NTData[:STATe]?
Factory Preset:
OFF
State Saved:
Saved in instrument state.
Dependencies and
Couplings:
The parameter is disabled if active mode is not Spectrum
Analyzer or a connection has not been established at the
specified IP address. Trace normalization cannot be turned ON
unless external source control is ON (that is, Amplitude state is
set to ON). If external source control is turned OFF while trace
normalization is ON, then trace normalization will be turned
OFF.
If a reference trace has not been stored prior to turning on trace
normalization, then trace normalization is turned on but an
error message is reported.
Turning on trace normalization changes the Y scale units to dB
when Preset is pressed. The trace data now displayed in trace 1
is the data measured in trace 1 minus the data in the reference
trace (trace 3).
Trace normalization will be turned OFF when the reference
trace display is set to View.
Example:
CALC:NDAT ON