Power Meter Operation
2
E4419B Power Meter User’s Guide
59
following key presses:
Press
,
. Select the appropriate softkey for the
measurement you require, , ,
,
, or
, then
or
.
11
The measurement result is now displayed.
Input Select
A B A/B B/A
A-B
B-A
N O T E
If the measurement frequency does not correspond directly to a frequency in the sensor
calibration table (if selected) and the frequency dependent offset table being used, the
power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table
or the frequency dependent offset table, the power meter uses the highest or lowest
frequency point in the appropriate table to set the calibration factor and offset.
The value of the calibration factors being used by the power meter to make a
measurement is displayed under the
or
softkeys.
A Cal Fac
B Cal Fac
Содержание E4419B
Страница 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Страница 14: ...xiv E4419B Power Meter User s Guide ...
Страница 22: ...xxii E4419B Power Meter User s Guide ...
Страница 44: ...2 Power Meter Operation 22 E4419B Power Meter User s Guide Contacting Agilent Technologies 109 ...
Страница 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Страница 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Страница 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Страница 175: ...153 User s Guide Agilent Technologies 4 Error Messages Introduction on page 154 Error Messages on page 156 ...
Страница 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...