Error Messages
4
User’s Guide
159
–230
Data corrupt or stale
This occurs when a
FETC?
is attempted and either a reset has been received or the power
meter state has changed such that the current measurement is invalidated (for example, a
change of frequency setting or triggering conditions).
–230
Data corrupt or stale;Please zero and calibrate Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A has
not been zeroed and calibrated, then any command which would normally return a
measurement result (for example
FETC?
,
READ?
, or
MEAS?
) will generate this error
message.
–230
Data corrupt or stale;Please zero Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A has
not been zeroed, then any command which would normally return a measurement result (for
example
FETC?
,
READ?
, or
MEAS?
) will generate this error message.
–230
Data corrupt or stale;Please calibrate Channel A
When
CAL[1|2]:RCAL
is set to
ON
and the sensor currently connected to channel A has
not been calibrated, then any command which would normally return a measurement result
(for example
FETC?
,
READ?
, or
MEAS?
) will generate this error message
–231
Data questionable;Lower window log error
This indicates that a difference measurement in the lower window has given a negative result
when the units of measurement were logarithmic.
–231
Data questionable;Upper window log error
This indicates that a difference measurement in the upper window has given a negative result
when the units of measurement were logarithmic.
–241
Hardware missing
The power meter is unable to execute the command because either no power sensor is
connected or it expects an Agilent E-Series or N8480 Series power sensor and one is not
connected.
–310
System error;Sensor EEPROM Read Failed - critical data not found or unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Sensor EEPROM Read Completed OK but optional data block(s) not found or
unreadable
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
–310
System error;Sensor EEPROM Read Failed - unknown EEPROM table format
This indicates a failure with your Agilent E-Series or N8480 Series power sensor. Refer to your
power sensor manual for details on returning it for repair.
Содержание E4419B
Страница 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Страница 14: ...xiv E4419B Power Meter User s Guide ...
Страница 22: ...xxii E4419B Power Meter User s Guide ...
Страница 44: ...2 Power Meter Operation 22 E4419B Power Meter User s Guide Contacting Agilent Technologies 109 ...
Страница 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Страница 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Страница 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Страница 175: ...153 User s Guide Agilent Technologies 4 Error Messages Introduction on page 154 Error Messages on page 156 ...
Страница 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...