Installation Note E4400-90392
17
Figure 4
Connector and Ribbon Cable Location (ESG-D Series Option 300)
Reassembling the Signal Generator
1. Refer to
. Replace the top cover (item 2), and torque the 11 screws
(item 1) that secure it to 9 in-lbs.
2. Refer to
. From the back, slide the instrument cover (item 6) onto the
signal generator.
3. Replace the four rear feet (item 4), and torque the four screws (item 5) to 21 in-lb.
4. Replace the two strap handles (item 1, one on each side of the instrument), and torque the
two screws (item 2) on each handle to 21 in-lb.
5. Replace the four bottom feet (item 3).