Appendix Service Information
Constant Voltage (CV) Verifications
179
Ser
vice
In
for
mat
ion
Constant Voltage (CV) Verifications
Constant Voltage Test Setup
If more than one meter or if a meter and an oscilloscope are used, connect
each to the (+) and (-) terminals by a separate pair of leads to avoid mutual
coupling effects. Use coaxial cable or shielded 2-wire cable to avoid noise
pick-up on the test leads.
Table A-3 Verification Programming Values
Voltage Programming and Readback Accuracy
This test verifies that the voltage programming and GPIB or RS-232 readback
functions are within specifications. Note that the readback values over the
remote interface should be identical to those displayed on the front panel.
You should program the power supply over the remote interface for this test
to avoid round off errors.
1
Turn off the power supply and connect a digital voltmeter between the (+) and
(-) terminals of the output to be tested as shown in Figure A-1.
2
Turn on the power supply. Select the output to be tested and the high voltage
range (20V/1.5A)
†
and then enable the output by sending the commands:
INST:SEL {OUT1|OUT2}
VOLT:RANG P20V
(E3646A model)
OUTP ON
3
Program the output voltage to 0 volt and current to full scale rated value
(1.5 A)
†
by sending the commands:
VOLT 0
CURR 1.5
(E3646A model)
4
Record the output voltage reading on the digital voltmeter (DVM). The reading
should be within the limit of (0 V ± 10 mV) for output1 or (0 V ± 25 mV) for
output2. Also, note that the
CV
,
Adrs
,
Limit
, and
Rmt
annunciators are on.
Model
Low voltage
range
High voltage
range
Model
Low voltage
range
High voltage
range
E3646A
8V/3A
20V/1.5A
E3647A
35V/0.8A
60V/0.5A
E3648A
8V/5A
20V/2.5A
E3649A
35V/1.4A
60V/0.8A
†
For E3646A model, and see
for other models.
Содержание E364XA Series
Страница 9: ...8...
Страница 15: ...14 Contents Contents...
Страница 16: ...1 Quick Start...
Страница 26: ...2 General Information...
Страница 38: ...3 Front Panel Operation and Features...
Страница 70: ...4 Remote Interface Reference...
Страница 122: ...5 Error Messages...
Страница 133: ...Chapter 5 Error Messages Calibration Errors 132...
Страница 134: ...6 Application Programs...
Страница 145: ...Chapter 6 Application Programs Example Program for Excel 97 144...
Страница 146: ...7 Tutorial...
Страница 157: ...Chapter 7 Tutorial Remote Programming 156...
Страница 158: ...8 Specifications...
Страница 165: ...Chapter 8 Specifications Supplemental Characteristics 164 Figure 8 2 Dimensions for Rack mounting...
Страница 166: ...Appendix Service Information...
Страница 175: ...Appendix Service Information General Disassembly 174 General Disassembly...
Страница 207: ......
Страница 208: ......
Страница 209: ......
Страница 210: ......
Страница 211: ......
Страница 212: ......
Страница 213: ......
Страница 214: ......
Страница 215: ......
Страница 216: ......
Страница 217: ......
Страница 218: ......