hpe1437_filter_setup
hpe1437_filter_setup
sets the digital filter bandwidth and decimation filter
parameters. This description also includes information on the following functions
which set or query the decimation filter parameters individually
hpe1437_filter_decimate
selects an extra factor of 2 decimation.
hpe1437_filter_decimate_get
gets current state of extra decimation
hpe1437_filter_bw
selects a signal filter bandwidth.
hpe1437_filter_bw_get
gets the signal filter bandwidth
VXI
plug&play
Syntax
#include “hpe1437.h”
ViStatus
hpe1437_filter_setup
(ViSession
id
, ViInt16
sigBw
, ViInt16
decimate
);
ViStatus
hpe1437_filter_decimate
(ViSession
id
, ViInt16
decimate
);
ViStatus
hpe1437_filter_decimate_get
(ViSession
id
, ViPInt16
decimatePtr
);
ViStatus
hpe1437_filter_bw
(ViSession
id
, ViInt16
sigBw
);
ViStatus
hpe1437_filter_bw_get
(ViSession
id
, ViPInt16
sigBwPtr
);
Parameters
id
is the VXI instrument session pointer returned by the
hpe1437_init
function.
sigBw
selects an alias protected signal filter bandwidth that is roughly
fs
/(2.56 *
2^(sigBw)) where
fs
is the ADC sample frequency. In zoom applications, where the
center frequency is generally not zero, the zoomfilter bandwidth is centered on the
frequency programmed with the
hpe1437_frequency_setup
function. For baseband
measurements the filter may equivalently be considered as a low pass filter of
approximately bandwidth
fs
/(2.56 * 2^(sigBw)) since the negative frequencies are
generally of no interest. The valid range of
sigBw
is 0 through 24. When
sigBw
= 0, no
digital filtering is applied to the signal and the module relies on the analog anti-alias
filter to limit the signal bandwidth to
fs
/2.56.
To more accurately calculate the bandwidth use the calculation
fs
* k/2^(sigBw) where:
k=.36 for .25 dB bandwidth
k=.44 for 3 dB bandwidth
k=.5 for 15 dB bandwidth
k=.62 for 110 dB bandwidth
For even more accuracy use the
hpe1437_filter_resp_get
function.
sigBwPtr
contains the current value of the
sigBw
parameter.
decimate
selects the data output sample rate. When this parameter is set to
HPE1437_OFF
the output sample rate is:
fs
when
sigBw
=0 or
fs
/2^(sigBw-1) when
sigBw
>0. When decimate is set to
HPE1437_ON
the output sample rate is reduced by
an additional factor of two by discarding alternate samples. You would normally want
to add the extra level of decimation in order to increase the displayed span.
E1437A User's Guide
VXI
plug&play
Programming Reference
4-28
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Содержание E1437A
Страница 7: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 47: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 123: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 177: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 178: ...6 Module Description 6 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 187: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 188: ...7 Verifying the E1437A 7 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 190: ...8 Replacing Assemblies 8 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 201: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 202: ...9 Backdating 9 1 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 207: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...
Страница 209: ...Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com...