To remove the cabinet
1.
Turn off the oscilloscope and disconnect the power cable.
2.
Rotate the handle to the up position as shown.
3.
Using the T20 TORX driver, remove the two screws from the rear of the cabinet.
4.
Using your thumbs, gently push on the rear-panel connectors to slide the oscilloscope out of the
cabinet.
Figure 1
Remove the cabinet.
T-20 Screw
T-20 Screw
Gently Push on the
connectors