![Agilent Technologies DSO6104A DSO6104L Скачать руководство пользователя страница 1](http://html.mh-extra.com/html/agilent-technologies/dso6104a-dso6104l/dso6104a-dso6104l_service-note_2868641001.webp)
P
P
P
R
R
R
I
II
O
O
O
R
R
R
I
II
T
TT
Y
Y
Y
S
S
S
A
A
A
F
FF
E
E
E
T
TT
Y
Y
Y
–
––
R
R
R
E
E
E
Q
Q
Q
U
U
U
I
II
R
R
R
E
E
E
S
S
S
I
II
M
M
M
M
M
M
E
E
E
D
D
D
I
II
A
A
A
T
TT
E
E
E
R
R
R
E
E
E
P
P
P
A
A
A
I
II
R
R
R
A
A
A
C
C
C
T
TT
I
II
O
O
O
N
N
N
A
A
A
N
N
N
D
D
D
S
S
S
P
P
P
E
E
E
C
C
C
I
II
A
A
A
L
LL
E
E
E
F
FF
F
FF
O
O
O
R
R
R
T
TT
T
TT
O
O
O
C
C
C
O
O
O
N
N
N
T
TT
A
A
A
C
C
C
T
TT
C
C
C
U
U
U
S
S
S
T
TT
O
O
O
M
M
M
E
E
E
R
R
R
S
S
S
ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
PRIORITY SAFETY
ACTION
STANDARDS:
CATEGORY:
X IMMEDIATELY LABOR: 0.5 Hours
LOCATION SERVICE [[]] RETURN USED [[]] RETURN
CATEGORY:
ON-SITE INVENTORY: [[]] SCRAP PARTS: [[]] SCRAP
X SERVICE CENTER X SEE TEXT X SEE TEXT
AVAILABILITY: ALWAYS AGILENT RESPONSIBLE UNTIL: ALWAYS
AUTHOR: DPM PRODUCT LINE: 1A
ADDITIONAL INFORMATION:
© AGILENT TECHNOLOGIES, INC. 2007
PRINTED IN U.S.A.
March 19, 2007
Page 1 of 9
MSO6104A-01A-S
S E R V I C E N O T E
Supersedes:
MSO6104A-01-S
MSO6104A 4+16-channel, 1 GHz Mixed Signal Oscilloscope
Serial Numbers: MY44001001
POSSIBLE SHOCK HAZARD
The loss of earth ground and mechanical tolerance variations in PCA and
instrument assembly may cause the power supply shield to short a trace to
ground creating a shock hazard.
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
P/N Description Qty.
54684-85402 Dielectric
insulator
1