Preparing the Gel-Dye Mix

1

Allow DNA dye concentrate (blue  ) and DNA gel matrix (red  ) to equilibrate to room temperature for 
30 min.

2

Add 25 μL of DNA dye concentrate (blue  ) to a DNA gel matrix vial (red  ).

3

Vortex solution well and spin down. Transfer to spin filter.

4

Centrifuge at 2240 g ± 20 % for 15 min. Protect solution from light. Store at 4 °C. 
Use prepared gel-dye mix within 6 weeks of preparation. 

Loading the Gel-Dye Mix

1

Allow the gel-dye mix to equilibrate to room temperature for 30 min before use.

2

Put a new DNA chip on the chip priming station.

3

Pipette 9 μL of gel-dye mix in the well marked 

.

4

Make sure that the plunger is positioned at 1 mL and then close the chip priming station.

5

Press plunger until it is held by the clip.

6

Wait for exactly 60 s then release clip.

7

Wait for 5 s, then slowly pull back the plunger to the 1 mL position.

8

Open the chip priming station and pipette 9 μL of gel-dye mix in the wells marked 

.

Loading the Marker

1

Pipette 5 μL of marker (green  ) in all  sample and ladder wells. Do not leave any wells 
empty.

Loading the Ladder and Samples

Technical Support

Please visit our support web page 

http://www.agilent.com/genomics/contactus

 to find information on your 

local Contact Center.

Further Information

Visit the 2100 Bioanalyzer site at 

http://www.agilent.com/genomics/bioanalyzer

. You can find useful 

information, support and current developments about the products and the technology.

1

Pipette 1 µL of DNA ladder (yellow  ) in the well marked 

.

2

In each of the 12 sample wells pipette 1 µL of sample (used wells) or 1 µL of de-ionized 
water (unused wells).

3

Put the chip horizontally in the adapter and vortex for 1 min at the indicated setting 
(2400 rpm).

4

Run the chip in the Agilent 2100 Bioanalyzer instrument within 5 min.