Agilent B2901/02/11/12A Precision Source/Measure Unit
Preparing the list data in advance
When you source some complicated waveform with the List Sweep Mode of the B2901/02/11/12A, it is convenient to
create the data list on a PC using the software of your choice and then transfer the data list file to the B2901/02/11/12A
via USB flash memory. TXT, PRN and CSV file format are available for data list file.
Step 1.
Develop CSV file which includes the following source list data.
(From 0 to 0.7 with 0.1 step, and from 0.71 to 1.30 with 0.01 step, and from 1.4 to 1.7 with 0.1 step)
Step 2.
Save CSV file as “listdata.csv” to the root folder of your USB flash memory device.
Step 3.
Attach the USB flash memory device which “listdata.csv” file is saved to the USB port on the front panel of the
B2901/02/11/12A.
Agilent B2900A Quick I/V Measurement Software is also available to edit the data list for List Sweep Mode. Please refer
to the Agilent Web Site below for the detailed information about Agilent B2900A Quick I/V Measurement Software.
http://www.agilent.com/find/b2900a