Agilent B2901/02/11/12A Precision Source/Measure Unit
Step 2.
Rotate
to select Channel 1 Trigger Parameters and set them up as below.
(Source Trigger Period 400 ms, Measurement Trigger Period: 400 ms)
Note) Source and Measurement Period should be the same as each other.
Now you’ve configured 400 ms trigger period. However, please note that FIXED current measurement range operation will
be used to control the trigger period strictly. The measurement range is selected by Limit value. In this example, 100 mA
measurement range will be used. If using AUTO measurement range operation is prior to controlling the trigger period
strictly, you may specify MANUAL trigger type with AUTO source and measurement trigger source by the following steps.
Step 3.
Press
to edit the trigger type, and then select
to set the trigger type to MANUAL.
Step 4.
Rotate
to select Channel 1 Trigger Parameters and set them up as below.
(Source Trigger Source: AUTO, Measurement Trigger Source: AUTO)
(2) Press MANUAL
(1) Rotate and press
Trigger Type
to edit