STATus Subsystem Introduction
l
Enable register cleared by:
l
STATus:Questionable:ENABle 0
l
l
Power cycle
l
does not clear enable register but it does clear event register.
l
does not affect this register.
l
The Query reads the enable register and returns a decimal value equal to the binary-weighted sum of all bits set in
the register.ter. For example, if bit 0 (value 1) and bit 1 ( value 2) are enabled, the query r3.
STATus:QUEStionable[:EVENt]?
Queries the event register for the
Questionable Data Register group
. This is a read-only register; the bits are cleared
when you read the register.
Parameter
Typical Return
(none)
+512
Read the event register (bit 9 set):
STAT:QUES?
l
Once a bit is set, it remains set until cleared by this query or
l
, and
have no effect on this register.
l
Query reads the event register and returns a decimal value equal to the binary-weighted sum of all bits set in the
register. For example, if bit 1 (value 2) and bit 9 ( value 512) are set, the query returns "+514".
290
Agilent 33500 Series Operating and Service Guide