Operational Limits
Tab. 10
INPUT VOLTAGE (V)
CONTROLLER STATUS
200 – 264
Operative with full performance
180 – 200
Operative with reduced performance
< 180 or > 264
Non operative condition
Status LED
Tab. 11
LED STATUS
PUMP STATUS
MEANING
Off
Stop
The pump is in stand-by and ready to start.
Green blinking
Ramp or
Autotuning
The pump has not yet reached its nominal
speed.
The pump is decreasing its speed due to a
gas load that needs power over the
available limit.
Green
Normal
The pump is running at its nominal speed.
Red
Fail
The pump is in fail. Check the error from
A-plus software and follow the actions
described in the table below.
Red blinking
Oil level warning
The oil level is under the limit. The led
stops the red blinking only when the oil
level is over the limit.
Orange
Starting self-test
At the start the pump is performing a self-
test for few seconds.
Orange blinking
Required maintenance The maintenance counter overcomes 8000
hours. To reset the counter, use the
window 358 from A-plus software.
Содержание 9499225
Страница 8: ...Contents MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 8 104...
Страница 33: ...Gebrauchsanleitung Entsorgung 2 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 33 104...
Страница 46: ...Mode d emploi Mise au rebut 3 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 46 104...
Страница 59: ...Instructions for Use Disposal 4 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 59 104...