![Agilent Technologies 9499225 Скачать руководство пользователя страница 91](http://html1.mh-extra.com/html/agilent-technologies/9499225/9499225_user-manual_2868747091.webp)
Technical Information
RS 232/RS 485 Communication Description
5
MS40+ Single Stage Rotary Vane Pumps User Manual / 87-901-008-01 (
H.00
)
91/104
Source: Pump
Destination: PC
02
80
06
03
38 35
STX ADDR ACK ETX CRC
Command: READ PUMP STATUS
Source: PC
Destination: Pump (with address = 3)
02
83
32 30 35 30 03 38 37
STX ADDR WINDOW RD ETX
CRC
Source: Pump (with address = 3 in stop status)
Destination: PC
02 83
32 30 35 30 30 30 30 30 30 03 38 37
STX ADDR
WINDOW
DATA (STATUS)
ETX
CRC
Window Meanings
Tab. 9
WIN
TYPE
R/W
DESCRIPTION
000
L
RW
START/STOP (1= START ; 0= STOP)
008
L
RW
REMOTE/SERIAL CONFIGURATION (1= Remote ; 0= Serial)
102
N
RW
SET POINT VALUE [Hz] (0 – 100)
105
N
RW
SET POINT HYSTERESIS [%] (0 – 100)
108
N
RW
BAUD RATE (0-4) [600, 1200, 2400, 4200, 9600]
120
N
RW
ROTATIONAL FREQUENCY SETTING [Hz] (40 - 60)
127
N
R
ROTATIONAL FREQUENCY SETTING [rpm] (read only)
200
N
R
BUS CURRENT [mA]
201
N
R
3PHASE VOLTAGE [Vrms]
202
N
R
POWER [W]
203
N
R
DRIVING FREQUENCY [Hz]
205
N
R
STATUS (0=stop; 1=wait interlock; 2=start; 3=autotuning; 5=normal; 6=fail)
Содержание 9499225
Страница 8: ...Contents MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 8 104...
Страница 33: ...Gebrauchsanleitung Entsorgung 2 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 33 104...
Страница 46: ...Mode d emploi Mise au rebut 3 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 46 104...
Страница 59: ...Instructions for Use Disposal 4 MS40 Single Stage Rotary Vane Pumps User Manual 87 901 008 01 H 00 59 104...