13 Maintaining the NPD
To Replace the NPD Bead Assembly
216
Maintaining Your GC
To Replace the NPD Bead Assembly
1
Gather the following:
•
Replacement NPD bead assembly, see
“Consumables and Parts for the NPD”
•
Lint-free gloves
•
T-10 Torx screwdriver
2
Set GC options for a new bead.
a
On the GC touchscreen, go to
Settings > Configuration > Detectors
to enable
Dry Bead
and
Auto Adjust
.
b
Then go to
Methods > Detectors
, scroll to
Detectors Specifics
, turn off the bead.
Agilent data system users: Make the same changes noted above, apply them to the GC,
and then save the data system method and shut down the instrument session.
3
Place GC in maintenance mode:
Maintenance > Instrument > Perform Maintenance >
Maintenance Mode > Start Maintenance.
Wait for the GC to become ready to continue (all
zones cooled to safe handling temperatures).
CAUTION
The bead is delicate. Be careful not to break or crack the bead. When performing
maintenance on the NPD, avoid touching the bead with your fingers, and prevent it from
coming in contact with other surfaces.
4
Turn off the NPD bead, and enable the
Dry Bead
and
Auto Adjust
settings.
Agilent data system users: After turning off the bead, save the data system method and
shut down the instrument session. (Note that in some data system versions, you may
need to use the GC touchscreen for these settings. To do this, the keypad must be
unlocked and you must close the GC parameters screen of the data system. Upload the
revised setting, then save the method and shut down the instrument session.)
5
Remove the GC detector top cover. See
“To Remove the Detector Top Cover”
WARNING
Hazardous voltages are present when the electronics top cover is open.
6
Remove the electronics cover. See
“To Remove the Electronics Cover”
7
Put on lint-free gloves before touching any of the detector parts.
WARNING
Be careful! The oven or detector fittings may be hot enough to cause burns.
CAUTION
Wear clean, lint-free gloves to prevent contamination of parts with dirt and skin oils.
8
Remove the screw from the NPD cover then lift up and set aside. See
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...