![Agilent Technologies 8890 Скачать руководство пользователя страница 18](http://html.mh-extra.com/html/agilent-technologies/8890/8890_maintaining-manual_2867678018.webp)
2
Removing Covers
To Remove the Detector Top Cover
18
Maintaining Your GC
To Remove the Detector Top Cover
This cover protects the detectors, valve box, and valve assembly. To remove the detector top
cover:
1
Turn off the GC and unplug the power cord.
2
Remove the detector cover by raising the cover up, then firmly lift up on the right side of the
cover to free it from the GC. Set the cover aside.
3
When finished with the maintenance procedure, place the detector cover vertically on the
GC and insert the left side first, then gently snap in the right side onto the GC.
WARNING
Be careful! With the power switch on, potentially dangerous voltages also exist:
• All electronics boards in the instrument.
• The internal wires and cables connected to these boards.
• The wires for the oven heater.
The covers shield these parts as the parts may be hot enough to cause burns. Never remove
a cover unless the instrument is unplugged.
CAUTION
Do not force the cover, either when installing it or closing it. This could break the plastic
parts.
Figure 2.
8890 GC covers
Detector cover
Pneumatics cover
Injection port cover
Left side cover
Power switch
Pneumatics cover
release button (on
both sides)
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...