Removing Covers
To Remove the Electronics Cover
20
Maintaining Your GC
To Remove the Electronics Cover
You may need to raise the electronics cover to perform NPD maintenance. The steps required
depend on whether or not an FPD+ is installed.
WARNING
Be careful! With the power switch on, potentially dangerous voltages also exist:
• All electronics boards in the instrument.
• The internal wires and cables connected to these boards.
• The wires for the oven heater.
The covers shield these parts as the parts may be hot enough to cause burns. Never remove
a cover unless the instrument is unplugged.
CAUTION
Raising the electronics cover exposes the GC electronics.
Without FPD+ installed:
1
Turn off the GC and unplug the power cord.
2
Raise or remove the detector top cover.
3
Loosen the captive screws on the right side cover, slide the cover backwards, then remove.
4
Loosen the captive screw located on the left side of the electronics cover.
5
Reach under the back of the electronics cover to release the clip holding it in place, then
raise the electronics cover and remove.
With FPD+ installed:
1
Turn off the GC and unplug the power cord.
2
Raise or remove the detector top cover.
Figure 3.
8890 GC electronics cover
Electronics cover
Loosen screw
Содержание 8890
Страница 1: ...Agilent 8890 Gas Chromatograph Maintaining Your Gas Chromatograph GC ...
Страница 8: ...8 Maintaining Your GC ...
Страница 22: ...2 Removing Covers To Remove the Electronics Cover 22 Maintaining Your GC ...
Страница 28: ...3 Maintaining the GC To Change the Hydrogen Sensor Calibration Gas Cylinder 28 Maintaining Your GC ...
Страница 48: ...4 Maintaining Capillary Columns To Remove the Ferrule From a CFT Fitting 48 Maintaining Your GC ...
Страница 120: ...7 Maintaining the COC Inlet To Bakeout Contaminants from the COC Inlet 120 Maintaining Your GC ...
Страница 142: ...8 Maintaining the MMI To Bakeout Contaminants from the MMI 142 Maintaining Your GC ...
Страница 178: ...10 Maintaining the VI To Bakeout Contaminants from the VI Inlet 178 Maintaining Your GC ...
Страница 208: ...12 Maintaining the TCD To Bakeout Contaminants from the TCD 208 Maintaining Your GC ...
Страница 240: ...14 Maintaining the ECD To Bakeout the ECD 240 Maintaining Your GC ...
Страница 264: ...16 Maintaining EPC Modules Installing or Replacing Frits in the PCM 264 Maintaining Your GC ...
Страница 276: ...17 Maintaining a Valve To Install the Upper Valve Box 276 Maintaining Your GC ...
Страница 282: ...www agilent com Agilent Technologies Inc 2019 First edition January 2019 G2790 90015 G3540 90015 ...