Technical Information
Interconnections – controller model 9698978M005
17
Turbo-V 1K-G User Manual
/
87-901-025-01 (A)
261/304
threshold: 500 Hz
hysteresis: 1%
activation type: "high level"
Figure 11
The set point output stays at 0 Vdc until the frequency becomes
higher than 505 Hz (that is 500 Hz + 1% of 500 Hz), then the output
goes at 24 Vdc and stays at 24 Vdc until the frequency becomes lower
than 495 Hz (that is 500 Hz – 1% of 500 Hz).
It is possible to delay the set point checking for a programmable
delay time.
The PROGRAMMABLE SET POINT signal has the following default
settings:
reference quantity: frequency
threshold: 643 Hz
hysteresis: 2%
activation type: high level
delay time: 0 second
Содержание 8698961R002
Страница 3: ...Turbo V 1K G Turbo V 1K G User Manual 87 901 025 01 A 3 304 Turbo V 1K G...
Страница 4: ...Turbo V 1K G 4 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 14: ...Contents 14 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 49: ...Mode d emploi Pr paration pour l Installation 3 Turbo V 1K G User Manual 87 901 025 01 A 49 304...
Страница 58: ...3 Mode d emploi Mise au Rebut 58 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 79: ...Manual de Istru es Prepara o para a instala o 5 Turbo V 1K G User Manual 87 901 025 01 A 79 304...
Страница 88: ...5 Manual de Istru es Elimina o 88 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 95: ...Bedrijfshandleiding Uitpakken 6 Turbo V 1K G User Manual 87 901 025 01 A 95 304...
Страница 104: ...6 Bedrijfshandleiding Afvalverwerking 104 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 290: ...17 Technical Information Purge Valve Installation 290 304 Turbo V 1K G User Manual 87 901 025 01 A...
Страница 299: ......
Страница 300: ......
Страница 301: ......
Страница 302: ......
Страница 303: ......