Technical Information
TV 401/301 MS Controller Installation
17
TV 401
/301 MS
User Manual / 87-901-013-01
(A)
245/
264
TV 401/301 MS Controller Installation
The 969-8979M005 controller can not be mounted on bottom of the
pump, like the standard controller, because it is not supplied with
the plastic pads for snap mounting. It can be mounted on any flat
surface by means of the three threaded holes on the controller casing
(see controller outline drawing).
Connection of the Fore-Vacuum Pump
A KF 16 flange is available to connect the TV 401/301 MS pump to the
fore-vacuum pump. A vacuum approved pipe can be used. If a rigid
pipe is used, any vibration generated by the mechanical pump must
be eliminated through the use of bellows.
NOTE
The TV 401/301 MS pump is characterized by its high compression ratio also for
oil vapors. When using a mechanical oil-sealed pump, it is advisable to install a
suitable trap between the turbopump and the fore-vacuum pump in order to
prevent oil backstreaming.
Содержание 869-8928R003
Страница 3: ...TV 401 301 MS TV 401 301 MS User Manual 87 901 013 01 A 3 264 TV 401 301 MS...
Страница 4: ...TV 401 301 MS 4 264 TV 401 301 MS User Manual 87 901 013 01 A...
Страница 224: ...17 Technical Information J5 In Out 224 264 TV 401 301 MS User Manual 87 901 013 01 A Figure 19 Case 1...
Страница 244: ...17 Technical Information Serial Cable Installation 244 264 TV 401 301 MS User Manual 87 901 013 01 A Figure 32...
Страница 258: ...17 Technical Information Accessories and Spare Parts 258 264 TV 401 301 MS User Manual 87 901 013 01 A...
Страница 259: ......
Страница 260: ......
Страница 261: ......
Страница 262: ......
Страница 263: ......