![Agilent Technologies 85103F Скачать руководство пользователя страница 31](http://html.mh-extra.com/html/agilent-technologies/85103f/85103f_installation-manual_2867932031.webp)
85103F Performance Upgrade Installation Manual
21
The 85103F Upgrade Package (8510A/B to 8510C)
Checking the 8510C
5. Turn off the 8510.
6. On the 85102:
a. Remove the top cover.
b. Disconnect W24 from the A17J4 Ref X connector.
c. Connect the BNC-to-SMB cable provided in the upgrade package from the rear panel
LO Phase Lock Out jack to the A17J4 Ref X connector.
7. Turn on the 8510.
8. Press
INSTRUMENT STATE RECALL
and select
[MORE], [FACTORY PRESET]
.
9. Press
PARAMETER
MENU
and select
[USER 1 a1]
.
10. Press
FORMAT
MENU
and select
[REAL]
.
11. Press
RESPONSE
AUTO
.
12. The signal displayed should ramp up as shown in
Figure 6-3
LO Phase Lock Out and Main Phase Lock Signals
13. On the 85102:
a. Remove the BNC-to-SMB cable.
b. Reconnect W24 to A17J4.
c. Leave the top cover off.