Environmental Requirements
Environmental Requirements
Temperature—What to Watch Out For
Changes in temperature can affect electrical characteristics. Therefore, the operating
temperature is a critical factor in performance. During a measurement calibration, the
temperature of the calibration devices must be stable and within the range shown in
IMPORTANT
Avoid unnecessary handling of the devices during calibration because your
fingers are a heat source.
Table 2-1
Environmental Requirements
Parameter Limits
Operating temperature
a
a. The temperature range over which the calibration standards maintain conformance to their
specifications.
+20
°
C to +26
°
C (+68
°
F to +79
°
F)
Error-corrected temperature range
b
b. The allowable network analyzer ambient temperature drift during measurement calibration
and during measurements when the network analyzer error correction is turned on. Also, the
range over which the network analyzer maintains its specified performance while correction
is turned on.
±
1
°
C of measurement calibration temperature
Storage temperature
−
40
°
C to +75
°
C (
−
40
°
F to +167
°
F)
Altitude
Operation
< 4,500 meters (
≈
15,000 feet)
Storage
< 15,000 meters (
≈
50,000 feet)
Relative humidity
Always non-condensing
Operation
0 to 80% (26
°
C maximum dry bulb)
Storage
0 to 90%
Содержание 85052C
Страница 5: ...1 1 1 General Information...
Страница 13: ...2 1 2 Specifications...
Страница 19: ...3 1 3 Use Maintenance and Care of the Devices...
Страница 27: ...85052C 3 9 Use Maintenance and Care of the Devices Gaging Connectors Figure 3 2 Gaging Male 3 5 mm Connectors...
Страница 33: ...4 1 4 Performance Verification...
Страница 36: ...4 4 85052C Performance Verification Recertification...
Страница 37: ...5 1 5 Troubleshooting...
Страница 41: ...6 1 6 Replaceable Parts...
Страница 45: ...A 1 A Standard Definitions...
Страница 62: ...Index 4 85052C Index part number 6 3 Z zeroing connector gage 3 8 3 10...